تخت احیا نوزاد
تخت احیا نوزاد تخت احیای نوزاد لیونا

Lyona l2100WPRO

The basic requirement for therapy and treatment of newborns, and in particular the ones with Low Birth Weight, is to have a temperature-controlled environment. To achieve this goal there are currently many methods which require usage of various devices, among which the most common, other than incubators, are the Infant Warmers. That is to say, open incubators with infrared radiant heating. These are most often used in delivery rooms and in intensive therapy centers. This system, which allows fast and easy access to the newborn, is the most used and efficient method for maintaining both the baby’s body and the bed at an ideal temperature.

TECHNICAL INFORMATION

Apgar timer (1-5-10 min)

Phototherapy duration timer

Electric controller with three modes: prewarm - manual - baby

Skin sensor to control the baby's body temperature in baby mod e

Easy access to the new born through the rotating capability of warmer to both sides

Cooling fan installed inside the warmer section for air conditioning

Two ceramic heaters to warm the bed and baby

Warmer guard to prevent the hand entry to the heater compartment

Protective foam to prevent damage to the user's head

X- Ray permeable mattress

Butterfly shields and gauges on both sides with polyamide plastic holders

Five failure alarms: over temperature, power failure, setting, checking and H/L

Holding handle in the main column for easy movement and install required accessories

Stand on both sides of the device to place the monitor and pulse oximeter

Trend/Rev.Trend, adjustment by means of gas spring

Fixed handle for Trend / Rev. Trend, adjustment

ABS cover for warmer and mattress platform

Tray for holding baby items on the base

Electrostatic powder coating

Four plastic swivel castors with 2 independent brakes in the front

Two oxygen cylinder holders

With IV pole holder on both sides

LED phototherapy lamps with a wavelength of 465 - 470 nano meter

Main power supply 220V - 50 Hz, 24 DC

Maximum power output 1200 VA

دارای نگهدارنده سرم خم دار

Light LED bulb

DIMENSIONS and BASICINFORMATION

Overall 1200 x 1060 x 1790 (mm)

Heater height up to mattress 750mm ± 10mm

Trend / Rev. Trend angle +15º / -15º

baby mode control range 34.5°C - 37.5°C

Angle of warmer module 0°, 30°, 60°, 90° for two ways

Accuracy of skin temperature sensor ± 0.1˚ C

Max load capacity 40 Kg

Bed net weight 110 Kg

Safe working load capacity 10 Kg

ENVIRONMENTAL TEMPERATURE :

Operating range 18°C - 30°C

STANDARDS

ISO 9001

ISO 13485

EN ISO 15223-1

EN ISO 14971

IEC 60601-1

IEC 60601-2-21

IEC 60601-2-50

IEC 62304

ACCESSORIES (Optional)

Two oxygen cylinder holders

Latest products

Click to buy Sina products
Consultation request
Catalog of products